EUTECH INSTRUMENTS CADMIUM EPOXY Instrukcja Obsługi Strona 59

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 68
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 58
...setting the standard again and again
®
57
Handheld Meters Multiparameter
Durable waterproof and dustproof
design with IP67 rating
Even with no probes attached and
the battery compartment open
Push-button calibration for all
parameters
6 points each for pH and
conductivity; 3 points for DO
Large backlit graphic display
Multi-line display with electrode
status indicator, calibration data,
and more
User-settable “calibration due”
alarm
Out-of-date or unperformed
calibrations are now things of
the past!
Built-in real-time clock
Time-and-date stamping meets
Good Laboratory Practice (GLP)
standards
Store up to 500 data sets
Infrared IrDA wireless technology
makes PC downloading convenient
and easy
Set point alarms
Audible warning when readings
are outside set points limit
Research-grade accuracy
Resolution to 0.001 pH and
accuracy to ±0.002 pH
Electrode status indicator
Calibration data provides electrode
diagnostic tool
Password protection
Security for calibration and
setup menus
Right: PD 650 meter
Left: Optional rubber
boot (included with
kit) provides added
protection
PD 650 pH/
Dissolved Oxygen Meter
Ideal for wastewater and aquaculture!
Specifications
Mode PD 650 meter
Range
pH –2.000 to 20.000 pH
Ion 0.001 to 19,900 ppm, molar
, or mg/L
mV ±2000 mV
DO 0.00 to 90.00 mg/L (ppm), 0.0 to 600.0%
Temperature –10.0 to 110.0°C (14.0 to 230.0°F), selectable
Resolution
pH 0.1/0.01/0.001 pH
Ion 2 or 3 digits
mV 0.1 mV
DO 0.01 mg/L (ppm), 0.1%
T
emperature 0.1°C (0.1°F)
Accuracy
pH ±0.002 pH
Ion ±0.5% full-scale (monovalent); ±1% full-scale (divalent)
mV ±0.2 mV
DO ±0.2 mg/L; ±2.0%
T
emperature ±0.5°C (0.9°F)
Calibration
pH
Up to 6 buffer values (select from 4 sets): USA: 1.68, 4.01, 7.01, 10.01, 12.45;
NIST
: 1.68, 4.01, 6.86, 9.18, 12.45; DIN: 1.09, 2.06, 4.65, 6.79, 9.23, 12.75;
or custom buffers
Ion
Up to 6 points
DO Two points; 100% and 0%
Temperature Offset 0.1°C (0.1°F) increments
DO salinity correction
Range: 0.0 to 50.0 ppt;
Resolution: 0.1 ppt
DO pressure correction
Range: 500 to 1499 mm Hg, 66.6 to 199.9 kPa
Resolution: 1 mm Hg, 0.1 kPa
Memory: up to 500 sets with GLP date and time
Output: infrared, IrDA
Real-time clock: time-and-date stamp on calibration
and stored data
Temperature compensation: automatic or manual
(selectable), from 0 to 100°C
Ambient operating temperature: 0 to 50°C (32 to 122°F)
Power: four 1.5 V AA batteries (included) or optional
universal AC adapter, up to 200 hours continuous use
Dimensions
Meter: 7.25"L x 3.25"W x 2.25"H
Boxed: 9.2"L x 9.2"W x 2.75"H (23 x 23 x 7 cm)
Weight
Meter: 1.0 lb (0.45 kg); Boxed: 2.0 lb (0.9 kg)
Ordering Information
Catalog number Description Included
WD-35432-02
PD 650 meter only Meter and batteries
WD-35432-00
PD 650 meter Meter, ”All-in-One“ electrode 35816-77, DO probe 35640-50, and batteries
WD-35432-70
PD 650 meter kit
Meter
, ”All-in-One“ electrode 35816-77, DO probe 35640-50, calibration
standards (60 mL each of pH 4.01, pH 7.00, electrode storage solution, and rinse
water), rubber boot, multiprobe holder
, batteries, and hard carrying case
WD-35816-77 “All-in-One” pH electrode with
built-in ATC; double junction, epoxy body,
10-ft cable with BNC connector
WD-35418-05 ATC probe. Use for temperature
compensation with any pH electrode without
built-in ATC
WD-35640-50 DO probe with 10-ft cable
WD-35434-85 Multiprobe holder. Holds one each
pH, conductivity, DO, and temperature probes
WD-35418-83 Optional adapter, 110/220 VAC
See pages 25–30 for more pH
electrodes; page 52 for DO probes.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682