...setting the standard again and again
®
14
pH/ORP Handheld Meters
Right: pH 110 meter
Left top: Easily
download data
readings into
your computer.
Left bottom: Use an
optional printer with
the pH 110 meter to
create hard copies of
your readings.
Applications
Industrial: Ideal for pH checks in water conditioning plants,
cooling towers, plating and finishing operations, food
processing water testing (for HACCP compliance), printing
and chemical industries, and water/wastewater treatment.
Environmental/Agricultural: Use in ecology studies,
aquariums, and hydroponics.
Laboratory: Use in all types of food processing,
environmental studies, chemical labs, titrations, and
quality assurance testing, especially where GLP data
management is required.
ORP: Use with an ORP electrode to monitor plating
operations, ozonation system efficiency, and other
redox measurements.
pH 110 Meter
Features RS-232 output and advanced setup mode!
Ordering Information
Cat. no. Description Included
WD-35615-22
pH 110 meter only Meter, two electrode holders, software, and batteries
WD-35615-20
pH 110 meter
Meter
, “All-in-One” pH/temperature probe 35811-71,
two electrode holders, software, and batteries
WD-35615-80
pH 110 meter kit
Meter
, “All-in-One“ probe, pH buffer pouches, three sample bottles, rinse
bottle, two electrode holders, software, hard carrying case, and batteries
Accessories
WD-35615-07 AC adapter, 110 VAC
WD-35615-08 AC adapter, 220 VAC
WD-35811-72 Double-junction
“All-in-One” electrode
WD-09376-00 Replacement batteries,
1.5 V AAA. Pack of 12
WD-35615-75 Portable meter carrying case.
Protects your instrument while still letting you
take measurements
RS-232 Computer/Printer Accessories
WD-35615-09 Computer cable, 9-pin male
to 9-pin female connects meter to your PC
WD-35622-00 Portable printer, 110 VAC,
rechargeable. Shpg wt 2 lb (0.9 kg)
WD-35622-05 Portable printer, 220 VAC,
rechargeable. Shpg wt 2 lb (0.9 kg)
WD-35622-59 Printer cable, 25-pin male
to 9-pin male connects meter to your printer
Temperature compensation:
manual or automatic (selectable) from 0 to 100°C
Operating temperature:
0 to 50°C (32 to 122°F)
Memory: stores up to 100 sets
RS-232 specifications
Baud rate: 2.4; 4.8; 9.6 and 19.2 kbps selectable
Stop bit: 1 or 2 selectable
Parity: odd (1); even (2); or none (0) selectable
RS-232 output connector: 9-pin female
Computer requirements: 386 and above that can run
Microsoft Windows
®
95 or higher, CD-ROM drive, hard
disk with approx. 500 KB free disk space, EGA monitor
or above, 9-pin serial port connecting cable, mouse
Power: four 1.5 V AAA batteries (included), >600 hours
continuous use; 9 V, 500 mA AC adapter (optional)
Dimensions
Meter: 7.5" x 3.5" x 1.75" (19 x 9 x 4.5 cm)
Boxed: 9.2" x 9.2" x 2.75" (23 x 23 x 7 cm)
Weight
Meter: 1.0 lb (0.5 kg); Boxed: 1.4 lb (0.6 kg)
Specifications
Mode pH mV Temperature
Range
–2.00 to 16.00 pH ±2000 mV (same for relative mV) –10 to 110.0°C (14 to 230ºF)
Resolution
0.01 and 0.1 pH (selectable) 0.1 mV ±399.9 mV, 1 mV beyond 0.1°C
Accuracy
±0.01 pH ±0.2 mV ±399.9 mV, 2 mV beyond ±0.5°C
Calibration
Up to 5 buffer values: 1.09,
1.68/3.06, 4.01/4.65, 6.79/6.86/
6.97/7.01, 9.18/10.01, 12.45/12.75
Offset to ±2000 mV
Offset 0.1°C increments
FREE Oakton
®
datalog software
on a CD-ROM
Helps organize data to import into
many popular spreadsheet programs
Five-point pH calibration
Choose from US, NIST or DIN
buffer sets
Automatic buffer recognition
For quick pH calibration
Features mV offset
For calibration to ORP standards,
zeroing the mV value, and pH
electrode diagnostics
Recall electrode slope, electrode
offset value, and previous
calibration data
Built-in memory function
Stores up to 100 pH, mV, or
relative mV readings, along with
temperature data
Selectable manual or automatic
temperature compensation (ATC)
For highest accuracy in any situation
°C/°F selectable
Hold and Auto-off functions;
Error messages
Power with batteries or
optional AC adapter
Use in the field or at your benchtop
using built-in meter stand
Available as a kit
Complete with calibration buffers,
rinse bottle, and protective carrying
case.
meter only
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682