EUTECH INSTRUMENTS CADMIUM EPOXY Instrukcja Obsługi Strona 18

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 68
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 17
...setting the standard again and again
®
16
pH/ORP Handheld Meters
Right: pH 610 meter
Left top: Easily download
data readings into your
computer.
Left bottom: Complete kits
available; includes meter,
carrying case and solutions.
pH 600 and pH 610 Meters
Expanded range, resolution, and accuracy—plus
increased memory and advanced communications!
Durable waterproof and dustproof
design with IP67 rating
Even with no probes attached and
the battery compartment open
Push-button pH calibration at up to
six points
Up to 15 buffer options with auto-
buffer recognition of USA, NIST,
DIN, and PWB standards
Large backlit graphic display
Multi-line display with electrode
status indicator, calibration data,
and more
User-settable “calibration due”
alarm
Out-of-date or unperformed
calibrations are now things of
the past!
Built-in real-time clock
Time-and-date stamping meets
Good Laboratory Practice (GLP)
standards
Store up to 500 data sets
Infrared IrDA wireless technology
makes PC downloading convenient
and easy
Set point alarms
Audible warning when readings
are outside set points limit
Research-grade accuracy
Resolution to 0.001 pH and accuracy
to ±0.002 pH on pH 610 model
Electrode status indicator
Calibration data provides electrode
diagnostic tool
Password protection
Security for calibration and setup
menus
Specifications
Mode pH 600 meter pH 610 meter
Range
pH –2.00 to 19.99 –2.000 to 19.999
mV ±2000 mV ±2000 mV
T
emperature –10.0 to 110.0°C (14.0 to 230.0°F), selectable –10.0 to 110.0°C (14.0 to 230.0°F), selectable
Resolution
pH 0.1/0.01 pH 0.1/0.01/0.001 pH
mV
0.1 mV 0.1 mV
T
emperature 0.1°C (0.1°F) 0.1°C (0.1°F)
Accuracy
pH ±0.01 pH ±0.002 pH
mV
±0.2 mV ±0.2 mV
T
emperature ±0.5°C (±0.9°F) ±0.5°C (±0.9°F)
Calibration
pH
Up to 6 buffer values (select from 4 sets): USA 1.68, 4.01, 7.01, 10.01, 12.45;
NIST: 1.68, 4.01, 6.86, 9.18, 12.45; DIN: 1.09, 2.06, 4.65, 6.79, 9.23, 12.75, or custom buffers
mV Offset up to ±150 mV
Temperature Offset 0.1°C (0.1°F) increments
Memory: up to 500 sets with GLP date and time
Output: infrared, IrDA
Real-time clock: time-and-date stamp on calibration
and stored data
Temperature compensation: automatic or manual
(selectable) from 0 to 100°C
Ambient operating temperature: 0 to 50°C (32 to 122°F)
Power: four 1.5 V AA batteries (included) or optional
universal AC adapter, up to 500 hours continuous use
Dimensions
Meter: 7.25"L x 3.25"W x 2.25"H
Boxed: 9.2"L x 9.2"W x 2.75"H (23 x 23 x 7 cm)
Weight
Meter: 1.0 lb (0.45 kg); Boxed: 2.0 lb (0.9 kg)
Ordering Information
Catalog number Description Included
WD-35418-02
pH 600 meter only Meter and batteries
WD-35418-12
pH 610 meter only Meter and batteries
WD-35418-00
pH 600 meter Meter, ”All-in-One“ probe 35816-71, and batteries
WD-35418-10
pH 610 meter Meter, ”All-in-One“ probe 35816-71, and batteries
WD-35418-70
pH 600 meter kit
Meter
, ”All-in-One“ probe, pH buffers (60 mL each of 4.01, 7.00, electrode
storage solution, and rinse water), rubber boot, two electrode holders,
batteries, and hard carrying case
WD-35418-80
pH 610 meter kit
Meter, ”All-in-One“ probe, pH buffers (60 mL each of 4.01, 7.00, electrode
storage solution, and rinse water), rubber boot, two electrode holders,
batteries, and hard carrying case
WD-35816-71 Replacement “All-in-One” pH/temperature probe, single-junction, epoxy body
WD-35816-72 “All-in-One” pH/temperature probe, double-junction, epoxy body
WD-35418-05 ATC probe. Use for temperature compensation with any pH electrode without built-in ATC
WD-35418-83 Optional adapter, 110/220 VAC
Multiparameter models available.
See pages 56–59.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
Przeglądanie stron 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 67 68

Komentarze do niniejszej Instrukcji