...setting the standard again and again
®
45
Benchtop Meters Conductivity
Dual display shows both
conductivity or TDS, and
temperature in °C or °F
Plus mode indicators and
annunciators to indicate meter status
Single- and multi-point automatic
or manual calibration
Customize meter calibration for
quick checks or for best meter
accuracy
Auto ranging across five ranges
Offers the best resolution for your
measurements
Advanced setup mode
Recalls previous calibration data,
conductivity/TDS effective cell
constants per range, and more
Built-in nonvolatile memory
Stores up to 50 conductivity or
TDS readings with corresponding
temperature
Automatic or manual temperature
compensation
Selectable cell constant
K = 0.1 for pure water; K = 1.0
for most applications; K = 10 for
concentrated solutions
Adjustable conductivity-to-TDS
conversion factor and
temperature coefficient
Gives highest accuracy in any
solution
Selectable normalization
temperature
Choose between 15 to 30°C
Built-in electrode holder arm;
selectable “Ready” indicator; Hold
function; diagnostic error messages
Includes conductivity cell
With cell constant of 1.0—ideal for
most applications
CON 510 Meter
An affordable way to take consistent, accurate
conductivity measurements in your laboratory!
WD-35608-50 Replacement conductivity cell,
Ultem body, stainless steel sensor, built-in
temperature sensor (K = 1.0). 3-ft cable
WD-35608-51 Conductivity cell, epoxy body,
platinum sensor, built-in temperature sensor
(K = 10). 3-ft cable
WD-35608-52 Conductivity cell, glass body,
platinum sensor, built-in temperature sensor
(K = 1.0). 3-ft cable
WD-35608-55 Conductivity cell, epoxy body,
platinum sensor, built-in temperature sensor
(K = 0.1). 3-ft cable
WD-35608-57 Conductivity cell, Ultem body,
stainless steel sensor, built-in temperature
sensor (K = 1.0). 10-ft cable
Specifications
Right: The CON 510 meter includes
electrode stand and handy slide-out instruction card.
Lower left: Conductivity/TDS/temperature probe has
removable probe guard for easy cleaning.
Ordering Information
Mode Conductivity TDS Temperature
Ranges
(Resolution)
0.00 to 20.00 μS (0.01 μS),
0.0 to 200.0 μS (0.1 μS),
0 to 2000 μS (1 μS),
0.00 to 20.00 mS (0.01 mS),
0.0 to 200.0 mS (0.1 mS)
0.00 to 10.00 ppm (0.01 ppm),
10.0 to 100.0 ppm (0.1 ppm),
100 to 1000 ppm (1 ppm),
1.00 to 10.00 ppt (0.01 ppt),
10.0 to 100.0 ppt (0.1 ppt),
100 to 200 ppt (1 ppt)
0.0 to 100.0°C (0.1°C),
32.0 to 212.0°F (0.1°F)
Accuracy
±1% full-scale or ±1 digit ±0.5°C, ±0.9°F
Calibration
Automatic: single or multi-point up to 4 points (84 μS, 1413 μS,
12.88 mS, 111.8 mS) Manual: single or multi-point up to 5 points
per range for multi-point calibration)
Offset 0.1°
increments
up to ±5°
T
emperature compensation:
automatic or manual
(selectable) from 0 to 100°C
Operating temp: 0 to 50°C (32 to 122°F)
Cell constant (K): 0.1, 1.0 or 10 selectable
Temp coefficient: adjustable from 0 to 10% per °C or °F
Conductivity-to-TDS factor: adjustable from 0.40 to 1.00
Normalization temp: adjustable from 15 to 30°C
Display: dual display of conductivity or TDS, with
temperature (°C or °F)
Error messages: diagnose operator error, electrode
error, and meter error. Pull-out instruction card
decodes message.
Memory: 50 sets
Power: 9 VDC, 500 mA; using AC adapter (included);
110 VAC, 60 Hz or 220 VAC, 50 Hz. AC adapter is
CSA and UL listed.
Dimensions
Meter: 9" x 7" x 2.375" (23 x 18 x 6 cm)
Boxed: 15.5" x 10" x 3.5" (39 x 25 x 9 cm)
Weight
Meter only: 1.25 lb (0.6 kg); Boxed 3 lb (1.4 kg)
Catalog number Description Includes
WD-35611-00
CON 510 meter
Meter, conductivity cell with an Ultem body/stainless steel sensor K = 1.0
(35608-50), AC adapter
, and electrode stand
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682