...setting the standard again and again
®
43
Handheld Meters Conductivity
Applications
Routine Testing: Ideal for quick, accurate conductivity
or TDS checks in labs, field, and schools.
Water Quality Testing: Use to test hard water, untreated
water, drinking water, incoming process water, and pool
water. Also useful in the hydroponics and agricultural
industries.
Industrial: Use to check pollution control, metal finishing,
cooling tower water, boiler water, drilling mud, brines, rinse
tanks, ponds, recirculating systems, and wastewater.
CON 400 and CON 410 Meters
Highly advanced microprocessor-based conductivity
meters with waterproof housing. . .and they even FLOAT!
Specifications
Ordering Information
Cat. no. Description Constant
WD-35608-50
Replacement conductivity cell,
Ultem body, steel sensor
.
3-ft cable
1.0
WD-35608-51
Conductivity cell, epoxy body,
platinum sensor
.
3-ft cable
10
WD-35608-55
Conductivity cell, epoxy body
,
platinum sensor
.
3-ft cable
0.1
WD-35608-57
Conductivity cell, Ultem body
,
steel sensor
. 16-ft cable
1.0
Conductivity Cells
Choose a cell with K = 0.1 for low range;
K = 1.0 for midrange; K = 10 for high range.
Feature a built-in temperature sensor;
measure 100 mm L x 12 mm dia.
Temperature compensation: automatic or manual
(selectable) from 0 to 50°C
Operating temperature: 0 to 50°C (32 to 122°F)
Temperature coefficient: adjustable from
0.0 to 10.0% per °C
Memory: up to 50 data sets
Power: four 1.5 V AAA batteries (included),
>100 hours continuous use
Dimensions
Meter: 7.5"L x 3.75"W x 2.25"H (19 x 9.5 x 5.7 cm)
Boxed: 9.2"L x 9.2"W x 2.75"H (23 x 23 x 7 cm)
Weight
Meter: 1.0 lb (0.5 kg); Boxed: 2.0 lb (0.9 kg)
CON 410
WD-35607-69 Calibration kit contains four each of “Singles” TDS calibration solution pouches
(447 μS, 1413 μS, 2764 μS, 15.0 mS), four rinse water pouches, one rinse bottle, three sample bottles,
all in a hard plastic carrying case with room for your meter and probe. Dimensions: 11.5"L x 14"W x 3"H.
Catalog number Description Included
WD-35608-00
CON 400 meter Meter, conductivity cell 35608-50 (K = 1.0), and batteries
WD-35608-10
CON 410 meter Meter, conductivity cell 35608-50 (K = 1.0), and batteries
Durable, waterproof and
dustproof IP67-rated design
Easy-to-clean housing
Ideal for dirty, wet environments and
food applications
Expanded memory
Stores and recalls up to 50 readings
with temperature, and previous
calibration data
Real-time clock
Stamps stored data and calibration
data with date and time; meets
standards for GLP
°C or °F selectable
Manual or automatic
temperature compensation
For the highest accuracy in any
situation
Selectable temperature coefficient
Adjustable from 0 to 10% per °C for
accurate compensation of almost
any solution
Selectable cell constant
K = 1.0 for general applications;
K = 0.1 for pure water;
K = 10 for concentrated solutions
CON 400 features:
Reads conductivity from
0 to 199.9 mS in five ranges
CON 410 also features:
Switch between conductivity
and TDS measurement with
a press of a button
Mode
Conductivity
(CON 400 and 410 meters)
TDS
(CON 410 meter only)
Temperature
(CON 400 and 410 meters)
Range/
Resolution
0.00 to 19.99 μS/0.01 μS,
0.0 to 199.9 μS/0.1 μS,
0 to 1999 μS/1 μS,
0.00 to 19.99 mS/0.01 mS,
0.0-199.9 mS/0.1 mS
0.00 to 9.99 ppm/0.01 ppm,
10.0 to 99.9 ppm/0.1 ppm,
100 to 999 ppm/1 ppm,
1.00 to 9.99 ppt/0.01 ppt,
10.0-99.9 ppt/0.1 ppt,
100-200 ppt/1 ppt
Epoxy-body cells:
0 to 80°C / 0.1°C
Glass-body cells:
0 to 100°C / 0.1°C
Accuracy
±1% full-scale ±1% full-scale ±0.5°C / ±0.5°F
Cell constant (K)
0.1, 1, 10.0 (selectable)
Conductivity-to-
TDS factor
0.4 to 1.0
Calibration
Up to five points
(one point per range)
Up to five points
(one point per range)
Offset 0.1° increments
meter only
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682