Instruction Manual PCD 650
65
3.10 DO Calibration in % Saturation Mode (with ATC)
The PCD 650 can be calibrated quickly and easily in air. In % Saturation, the
meter is able to perform either a one point calibration or a 2 point calibration.
For one point calibration, it is recommended that you perform a 100%
Saturation calibration in saturated air. If you opt for 2 point calibration, you
can calibrate for 100% Saturation in saturated air and 0% Saturation using a
zero oxygen solution.
The PCD 650 meter is capable of measuring barometric pressure with its
built-in pressure sensor. In the event, the pressure reading is inaccurate, you
can calibrate the value from the setup menu. For setting barometric pressure
range, please refer page 86.
3.10.1 To calibrate 100% saturation
1. Switch on the meter and make sure the meter is in % saturation mode.
2. Hold the probe in the air gently with the sensor facing down and press CAL
(F2) to start calibration.
3. The meter shows ‘Dissolved O
2
Calibration-Rinse Electrode’ screen [Figure
48-(a)] for few seconds to prompt user to rinse electrode with de-ionized
water before calibration
(a) (b)
Figure 58: Saturation Calibration Screens
4. Rinse the probe well with de-ionized water. For best result, blot the end of
the probe dry. Do not touch the membrane.
5. The meter then shows saturation calibration screen [Figure 58-(b)]. The
primary display will show the current value of measurement. Wait for the
reading to stabilize.
Note: If the meter is password protected, you will be prompted to enter a
password. Refer ‘Accessing Calibration mode when password protection
enabled’ in page 42.