
Instruction Manual PCD 650
18
2.2.5 Automatic Temperature Compensation (ATC)
Connect an appropriate probe with built in ATC probe to the meter and select
‘ATC mode’ in the temperature setup (See page 98) for the
pH/conductivity/DO reading to be automatically compensated for temperature
variations.
If you select ‘ATC’ without connecting a probe to the meter, the LCD shows
‘UNDER’ for temperature reading.
Figure 18: Under range
NOTE: The factory default value for normalization temperature is 25 °C.
2.2.6 Manual Temperature Compensation (MTC)
If ATC probe is not available, you can choose to manually compensate for
temperature. This is suitable when the temperature of your sample is
sufficiently stable. Select ‘MTC mode’ in the temperature setup (see page 98).
Press CAL (F2) and then press NEXT key. Press TEMP (F1) to go to
temperature calibration. Enter the temperature value of your sample. See
page 45 for more details on temperature calibration.
2.2.7 Alarm set point (For pH/conductivity/DO)
You can set the meter to display an alarm when the
pH/conductivity/salinity/resistivity/TDS/DO(%)/DO(mg/L) reading goes higher
or lower than predefined set points. Refer Setup of respective modes to set
alarm points for pH, conductivity and DO.
Figure 19 : Hi alarm
2.2.8 Calibration Due (CAL-DUE) Indicator (For pH/conductivity/DO)
You can set a reminder to be displayed in the LCD when the next calibration
is due. Set the number of days in the pH/conductivity/DO Setup and the
meter will remind you when the days elapse from your last calibration date.
2.2.9 Resolution of pH Reading
By default, the resolution is set to 0.00. You can change the resolution in the
setup mode. pH resolution can be adjusted from 0.1 to 0.001 pH (See page
83)
2.2.10 pH buffer group for calibration and auto-recognition
This option allows you to choose buffer group from four internationally
recognized standard buffer groups namely USA, NIST, DIN & PWB. The
default buffer group is USA. Refer ‘pH Setup’ section in page 82 for
information on how to set the meter to a particular buffer group.
Komentarze do niniejszej Instrukcji