Instruction Manual PCD 650
39
3. Calibration Mode
3.1 About Calibration
PCD 650 series meters are factory calibrated and allows you to measure
pH/mV/ion/conductivity/resistivity/TDS/salinity/DO(%)/DO(mg/L). Calibrate to
all measurement ranges to ensure the highest accuracy throughout all
measurement range. However, it is recommended that, for higher accuracy,
you calibrate your meter before you make measurements for the first time.
Calibration should be carried out each time a new electrode is attached to the
meter or when you suspect that the meter/electrode is out of calibration.
3.1.1 About Temperature Calibration
It is important to ensure that temperature calibration is carried out prior to pH,
conductivity and DO calibration since temperature readings affect the
accuracy of pH, conductivity and DO measurements. You need to perform
temperature calibration only if the temperature value displayed on the meter
is different from that of a calibrated thermometer or if cell constant setting is
changed. A temperature offset calibration of ± 5 °C/± 9 °F from the default
reading is allowed for ATC mode.
3.1.2 About pH Calibration
When you re-calibrate your meter, previous pH calibration points are
replaced on a point by point basis. For example, if you previously calibrated
your meter at pH 4.01, 7.00, and 10.01, and you have now re-calibrated at
pH 7.00, the meter retains the old calibration data at pH 4.01 and pH 10.01.
The meter shows previously calibrated points in the display when the meter is
in pH measurement mode. To completely re-calibrate the meter, or when you
use a replacement probe, it is best to clear the previous calibration and re-
calibrate the meter at all points. Refer page 46 for information on how to
clear previous pH calibration.
The meter supports four internationally recognized standard buffer groups:
USA, NIST, DIN & PWB. The default buffer group is USA. Refer page 82 for
information on how to set the meter to a particular buffer group
The meter is capable of calibrating up to 6 points, depending on the standard
buffer selected. When completely re-calibrating the meter, the first point of
calibration should be one of the following, depending on your choice of
standard buffer group.
Table 2: Buffer Groups and Calibration Points
Buffer
Group
First Point of Calibration
(when complete re-calibration)
Other Available Calibration Points
USA pH 7.00 pH 1.68, 4.01, 10.01, and 12.45
NIST pH 6.86 pH 4.01, 6.86, 9.18, and 12.45
DIN pH 6.79 pH 1.09, 3.06, 4.65, 9.23, and 12.75
PWB pH 6.97 pH 4.10