8. Rinse the probe with de-ionized water. Place it in the next solution and follow
the steps 2 to 7 to calibrate other points, if ‘MULTI’ point calibration is selected.
9. Press ESC (F4) to go to measurement mode. The meter shows the salinity
reading of the solution in measurement mode.
3.8.1 Salinity Calibration Report
Calibration report gives you information on calibration. It includes date & time on
which the last calibration was done and the offset.
To View salinity Report:
1. From salinity measurement mode, press left
or right arrow key to
navigate to other available functions until you see REPO function in the LCD
2. Press REPO (F2) key. The calibration report is shown in the display [Figure
55].
3. Press PRIN (F2) to transfer the calibration report to the computer.(Refer page
37 to print data)
Figure 55 : Salinity report screen
Note: The function of the keys available on the salinity cell
constant/calibration/ report screen is same as previously described in
conductivity calibration.