Instruction Manual TN-100/ T-100
21
9 ADDENDUM 2: GUIDE TO GOOD MEASUREMENT
TECHNIQUE
Turbidity is a very complex analytical measurement which can be affected
by many factors. Some are inherent in the instrument’s design such as
angle of detection, light beam aperture, incident beam wavelength and
colour sensitivity of the photocell.
However, there are other factors such as stray light, air bubbles and care
of vial, which can be prevented through proper care of equipment and
accessories, and in the operating procedure for measurement. Here are
some points you may want to note:
9.1
Maintain sample vials in good condition
Sample vials must be meticulously clean and free from significant
scratches. It should be treated on the outside with a thin coat of silicone
oil. This is to mask minor imperfections and scratches that may contribute
to stray light. Sample vials should be handled only by the top to avoid dirt
accumulation (or deposits) and fingerprints that might interfere with the
light path.
More information is found in Section 5.1 - Vials – Handling, Cleaning and
Care on page 16.
9.2
Match Sample Vials
Best accuracy and repeatability of turbidity measurement are achieved
using a single, indexed vial. However, for more convenience, different
vials can be used for measurement provided their readings with the same
solution are matched. That is, the meter gives identical readings or within
the specified repeatability and accuracy of the meter.
Select a few vials. After the sample vials are cleaned, fill them with ultra-
low turbidity water. Allow the sample vials to stand and for air bubbles to
rise. Polish sample vials with silicone oil and take turbidity measurement at
several points while rotating it in the sample well. Find the position where
turbidity reading is the lowest and index it for each vial. Whenever these
sample vials are used, use the indexed mark to position each vial into the
sample well. Choose those vials that match the readings.
NOTE:
Not all vials can be matched because of some
manufacturing variations.
9.3 Degassing
Air or other trapped gases should be removed before measurement.