Instruction Manual TN-100/ T-100
10
3 TURBIDITY MEASUREMENT
3.1 General Information
The waterproof TN-100/ T-100 turbidimeter allows you to measure the
turbidity of a grab sample. The turbidity is reported in Nephelometric
Turbidity Units (NTU). Readings above 1000 NTU are outside the range of
this instrument.
NOTE:
Before switching on the TN-100/ T-100 turbidimeter,
a sample vial MUST be placed in the sample well.
You can use any of the calibration standards for this
purpose.
The meter will go through a power sequence as shown in Figure 6.
CAL 13 CAL 24
mg/l
NTU
ppm
%pH
TAvg
NTU
OFF
ON
"Rd" will flash 10 times
before a reading appears
(assuming you have
placed a vial into the
sample well of the meter
A reading will only appear
if you have placed a vial
or any of the calibration
standards provided into
the sample well.
Otherwise an error
message "Err 8" will
appear.
Figure 6: Power-up sequence
3.2
Turbidity Measurement Procedure
An accurate turbidity measurement depends on good measurement
techniques. Factors such as clean sample vials, positioning of vial in the
sample well, meter calibration, handling of meter, and others, have to be
taken into consideration. Please see Section 5.1: Vials – Handling,
Cleaning and Care on page 16 and Section 9: ADDENDUM 2: Guide to
Good Measurement Technique on page 21 for more information.