Instruction Manual TN-100/ T-100
20
Figure 14 shows the waterproof TN-100/ T-100 turbidimeter basic optical
system. It includes a light source and a detector to monitor the light
scattered at 90° with respect to the incident beam.
90°
Lens
Glass Sample Vial
Transmitted
Light
Scattered Light
Lamp
Figure 14: Basic Nephelometric arrangement for turbidity measurements
8.4
Nephelometric Turbidity Units (NTU)
Nephelometric Turbidity Units (NTU): Unit of measure used when relating
the light scattered by a liquid media to the light scattered by a known
concentration of a standard solution. This unit of measure is recognized as
a measure of the optical clarity of an aqueous sample. NTU is the
accepted unit of measurement for turbidity.
Another unit commonly used to measure turbidity is Formazin Turbidity
Unit (FTU). The two units of measure of turbidity are equivalent: 1 NTU = 1
FTU.
8.5
Indexing a Vial
The United States Environmental Protection Agency (US EPA)
recommends that vials used for turbidimeter calibration or sample
measurement be indexed.
To index a sample vial, slowly rotate the vial throughout one complete
revolution (360°). While rotating the sample vial, observe the display and
locate the position that the vial is in which provides the lowest turbidity
reading. This position is the indexed position of the vial.
Mark this position on the vial (not on the cap) against the mark on the
meter.
After indexing a vial, make sure the vial will always be placed inside the
sample well in the indexed position.