CON 700 Conductivity/ºC/ºF Bench Meter
Technolo
g
y
M
ade
E
a
s
y
.
.
.
Part of Thermo Fisher Scientific 68X541702 Rev. 1 Jan 2010
Przeglądanie stron 0
1 2 3 4 5 6 ... 23 24

Podsumowanie treści

Strona 1 - Instruction Manual

Instruction Manual CON 700 Conductivity/ºC/ºF Bench Meter TechnologyMadeEasy... Part of Thermo Fisher Scientific

Strona 2

The factory default is Single-Point Calibration. See Section 6.11 to change this setting. 4.3 General Calibration Tips For best results, periodic cal

Strona 3 - TABLE OF CONTENTS

4.5 Manual Conductivity & TDS Calibration Procedure 1) Press as needed to select conductivity (μS or mS) or TDS (ppm or ppt) calibration. 2) D

Strona 4

2) Dip the probe into the sample beyond the upper steel band (utilize the fill line on the outside of the probe guard for reference). 3) Allow time

Strona 5

2) To recall data from memory, press . The location of the most recent stored data is displayed first. Press or to select the location of the des

Strona 6 - GETTING STARTED

Press to confirm. 6.5 P3. (C hrenheit) 2 ºC ºF elcius Or FaPress . Press or to select ºC or ºF. Press to3 AtC (Automatic Tem confirm. 6.6

Strona 7

Press . Press or to select the desired value (0.00 to 10.0). to confirm. Press 6.9 P3.6 t.nr (Normalization Temperature in ºC) When Au are adj

Strona 8

12 performa ce in extreme sample ranges. Use this setup function to eter default is 1.0 to match the nchange the cell constant if necessary. Minclud

Strona 9

13 7. CALCULATING TDS CONVERSION FACTOR You can calibrate TDS using the value of the calibration standard solution at a standard temperature such a

Strona 10

8. CALCULATING TEMPERATURE COEFFICIENTS To determine the temperature coefficient of your sample solution use this formula: Where: tc = Temperatur

Strona 11 - 5. MEASUREMENT

15 9. REPLACEMENTS AND ACCESSORIES Part number Ordering Code Item Description Eutech Instruments Oakton Instruments CON 700 Benchtop with electrod

Strona 13 - 6. SETUP FUNCTIONS

10. TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION No display Main power not switched on. AC Adapter socket not inserted properly. a) Switch on the p

Strona 14

17 11. SPECIFICATIONS Conductivity Range 0 to 20.00, 200.0, 2000 µS 0 to 20.00, 200.0 mS TDS Range (0.5 TDS factor) 0 to 10.00, 100.0, 1000 ppm

Strona 15

18 12. WARRANTY This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from

Strona 17

For more information on our products, please contact our channel partner or visit our websites listed below: Oakton Instruments 625 E Bunker

Strona 18

TABLE OF CONTENTS 1. INTRODUCTION ... 1 2. GETTING STARTED...

Strona 19

6.12 CELL (Nominal Cell Constant) ...11 6.13 P4.0 rSt (Reset) ...

Strona 20 - 10. TROUBLESHOOTING GUIDE

1 1. INTRODUCTION Thank you for purchasing our Con 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The

Strona 21 - 11. SPECIFICATIONS

2. GETTING STARTED 2.1 Keypad Functions Powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used.

Strona 22

2.2 LCD Annunciators 2.3 Meter Connections CON/Temp 8-pin DIN connection for 2-cell Con/TDS/Temp electrode DC Power supply 3

Strona 23

3. ELECTRODE INFORMATION The CON700 includes an electrode with a nominal cell constant of k = 1.0, built-in temperature sensor, and 1 meter cable. Th

Strona 24

5 The factory default is automatic conductivity calibration. See Section 6.10 to change this setting. There is no automatic calibration value availa