Press to confirm.
6.5 P3. (C hrenheit) 2 ºC ºF elcius Or Fa
Press
.
Press or to select ºC or ºF.
Press to
3 AtC (Automatic Tem
confirm.
6.6 P3. perature Compensation)
Press
.
Press or to select “Yes” or “No”.
Press
to confirm.
6.7 P3. ot Solids factor)
ress
4 tdS (T al Dissolved
P
.
Press or to select the desired TDS factor (.40 to
1.00).
Press
to confirm.
6.8 P3.5 t.CO (Temperature Coefficient)
T
he temperature coefficie ount of change in conductivity per
0 is factory set to a
ost applications this will
See Section 8 – Calculating Temperature Coefficients.
nt is the am
degree temperature (% per ºC). The CON 70
temperature coefficient of 2.1 % per ºC. For m
provide good results. The meter allows adjustment from 0.0 to 10.0.
TIP: Select 0.0% for uncompensated measurements. The temperature
will be measured by the electrode and displayed in measurement
mode—without compensation.
10