Strona 1 - TN-100/ T-100

Instruction Manual TN-100/ T-100 Portable Turbidimeter Tech n o logyMadeEasy...68X357701 Rev. 0.1 05/03

Strona 2

Instruction Manual TN-100/ T-100 62.3 Calibration Procedure 1. Place TN-100/ T-100 turbidimeter on a flat and level surface. 2. Select the c

Strona 3 - Rev. 0.1 05/03

Instruction Manual TN-100/ T-100 7CAL 1800 NTUNTUCAL 1NTUNTUCAL 3CAL 3NTUCAL 2NTUCAL 2READENTERNTUCAL 1CALPlace the CAL 1 standard into the samp

Strona 4 - TABLE OF CONTENTS

Instruction Manual TN-100/ T-100 82.4 Restoring Factory Calibration The TN-100/ T-100 turbidimeter allows you to reset the meter back to the fa

Strona 5 - 1 OVERVIEW

Instruction Manual TN-100/ T-100 9 READENTERREADENTERPress and keys simultaneously and hold for 2 seconds before releasing UP ON/OFF [-Rd-]

Strona 6 - CAL 13 CAL 24

Instruction Manual TN-100/ T-100 103 TURBIDITY MEASUREMENT 3.1 General Information The waterproof TN-100/ T-100 turbidimeter allows you to meas

Strona 7 - WATERPROOF

Instruction Manual TN-100/ T-100 113.2.1 Preparation of Sample Vial 1. Grab liquid with the plastic bottle provided. 2. Obtain a clean and d

Strona 8 - Battery cover

Instruction Manual TN-100/ T-100 123.2.2 Measurement Procedure 1. Place TN-100/ T-100 turbidimeter on a flat and level surface. 2. Place the

Strona 9 - 2 TURBIDITY CALIBRATION

Instruction Manual TN-100/ T-100 13NOTE: 1. NEVER pour liquid directly into the sample well of the instrument. Always use a vial. The instrume

Strona 10 - Calibration Procedure

Instruction Manual TN-100/ T-100 14 For continuous measurement: 1. Make sure the meter is sitting on a flat and level surface and is in measure

Strona 11 - 0.02 NTU

Instruction Manual TN-100/ T-100 154 TROUBLESHOOTING GUIDE The TN-100/ T-100 turbidimeter routinely performs self-diagnostics, and will automa

Strona 13 - METER OFF

Instruction Manual TN-100/ T-100 165 ROUTINE MAINTENANCE The supplied carrying case is optimal for protecting the instrument. If you do not pla

Strona 14 - 3 TURBIDITY MEASUREMENT

Instruction Manual TN-100/ T-100 176 ACCESSORIES The items shown below are recommended accessories and replacement parts for the instrument. I

Strona 15

Instruction Manual TN-100/ T-100 187 SPECIFICATIONS MODEL NO. TN-100/ T-100 Measurement Method ISO 7027 compliant nephelometric method (90°) Me

Strona 16 - Figure 11: Reading turbidity

Instruction Manual TN-100/ T-100 198 ADDENDUM 1: TURBIDITY 8.1 Definition Turbidity is defined as an “expression of the optical property that

Strona 17 - Unknown

Instruction Manual TN-100/ T-100 20Figure 14 shows the waterproof TN-100/ T-100 turbidimeter basic optical system. It includes a light source an

Strona 18 - Figure 13: Continuous

Instruction Manual TN-100/ T-100 219 ADDENDUM 2: GUIDE TO GOOD MEASUREMENT TECHNIQUE Turbidity is a very complex analytical measurement which ca

Strona 19 - 4 TROUBLESHOOTING GUIDE

Instruction Manual TN-100/ T-100 22Degassing is recommended even if no bubbles are visible. There are three methods commonly used for degassing:

Strona 20 - 5 ROUTINE MAINTENANCE

Instruction Manual TN-100/ T-100 239.6 Calibration 1. Do not open the vials with calibration standards. 2. Check that the standards have not

Strona 21 - 6 ACCESSORIES

Instruction Manual TN-100/ T-100 2410 WARRANTY The TN-100/ T-100 meter is supplied with a 2-year warranty from manufacturing defects and calibra

Strona 22 - 7 SPECIFICATIONS

Instruction Manual TN-100/ T-100 25CALIBRATION CHECKLIST Month ____________ Year ___________ Date Initials Value Standard Comments

Strona 23 - 8 ADDENDUM 1: TURBIDITY

Preface This manual functions in two ways: first as a step by step guide to help you operate the waterproof TN-100/ T-100 turbidimeter; second, it ser

Strona 26 - Timeliness of Sample

For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website

Strona 27 - T = Td * (Vs + Vd) / Vs

TABLE OF CONTENTS 1 OVERVIEW... 1 1.1 Unpacking TN-100/ T-100 M

Strona 28 - 11 RETURN OF ITEMS

Instruction Manual TN-100/ T-100 11 OVERVIEW Thank you for selecting the waterproof portable TN-100/ T-100 turbidimeter. The TN-100/ T-100 allow

Strona 29 - CALIBRATION CHECKLIST

Instruction Manual TN-100/ T-100 2Sample VialSample WellIndex MarkLiquid Crystal DisplayKeypad Figure 1: Parts of TN-100/ T-100 Turbidimeter 1.2

Strona 30

Instruction Manual TN-100/ T-100 31.3 Keys and Functions The keypad has five keys: ON/OFF, CAL, ▲, ▼, and READ/ENTER. Keys Functions ON/OFF -

Strona 31

Instruction Manual TN-100/ T-100 41.4 Battery Installation Four AAA-sized batteries are included in your meter’s packaging: 1. Use a Philips

Strona 32

Instruction Manual TN-100/ T-100 52 TURBIDITY CALIBRATION The TN-100/ T-100 turbidimeter was calibrated and tested prior to leaving the factory.