Instruction Manual TN-100/ T-100 62.3 Calibration Procedure 1. Place TN-100/ T-100 turbidimeter on a flat and level surface. 2. Select the c
Instruction Manual TN-100/ T-100 7CAL 1800 NTUNTUCAL 1NTUNTUCAL 3CAL 3NTUCAL 2NTUCAL 2READENTERNTUCAL 1CALPlace the CAL 1 standard into the samp
Instruction Manual TN-100/ T-100 82.4 Restoring Factory Calibration The TN-100/ T-100 turbidimeter allows you to reset the meter back to the fa
Instruction Manual TN-100/ T-100 9 READENTERREADENTERPress and keys simultaneously and hold for 2 seconds before releasing UP ON/OFF [-Rd-]
Instruction Manual TN-100/ T-100 103 TURBIDITY MEASUREMENT 3.1 General Information The waterproof TN-100/ T-100 turbidimeter allows you to meas
Instruction Manual TN-100/ T-100 113.2.1 Preparation of Sample Vial 1. Grab liquid with the plastic bottle provided. 2. Obtain a clean and d
Instruction Manual TN-100/ T-100 123.2.2 Measurement Procedure 1. Place TN-100/ T-100 turbidimeter on a flat and level surface. 2. Place the
Instruction Manual TN-100/ T-100 13NOTE: 1. NEVER pour liquid directly into the sample well of the instrument. Always use a vial. The instrume
Instruction Manual TN-100/ T-100 14 For continuous measurement: 1. Make sure the meter is sitting on a flat and level surface and is in measure
Instruction Manual TN-100/ T-100 154 TROUBLESHOOTING GUIDE The TN-100/ T-100 turbidimeter routinely performs self-diagnostics, and will automa
Instruction Manual TN-100/ T-100 165 ROUTINE MAINTENANCE The supplied carrying case is optimal for protecting the instrument. If you do not pla
Instruction Manual TN-100/ T-100 176 ACCESSORIES The items shown below are recommended accessories and replacement parts for the instrument. I
Instruction Manual TN-100/ T-100 187 SPECIFICATIONS MODEL NO. TN-100/ T-100 Measurement Method ISO 7027 compliant nephelometric method (90°) Me
Instruction Manual TN-100/ T-100 198 ADDENDUM 1: TURBIDITY 8.1 Definition Turbidity is defined as an “expression of the optical property that
Instruction Manual TN-100/ T-100 20Figure 14 shows the waterproof TN-100/ T-100 turbidimeter basic optical system. It includes a light source an
Instruction Manual TN-100/ T-100 219 ADDENDUM 2: GUIDE TO GOOD MEASUREMENT TECHNIQUE Turbidity is a very complex analytical measurement which ca
Instruction Manual TN-100/ T-100 22Degassing is recommended even if no bubbles are visible. There are three methods commonly used for degassing:
Instruction Manual TN-100/ T-100 239.6 Calibration 1. Do not open the vials with calibration standards. 2. Check that the standards have not
Instruction Manual TN-100/ T-100 2410 WARRANTY The TN-100/ T-100 meter is supplied with a 2-year warranty from manufacturing defects and calibra
Instruction Manual TN-100/ T-100 25CALIBRATION CHECKLIST Month ____________ Year ___________ Date Initials Value Standard Comments
Preface This manual functions in two ways: first as a step by step guide to help you operate the waterproof TN-100/ T-100 turbidimeter; second, it ser
For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visit our website
TABLE OF CONTENTS 1 OVERVIEW... 1 1.1 Unpacking TN-100/ T-100 M
Instruction Manual TN-100/ T-100 11 OVERVIEW Thank you for selecting the waterproof portable TN-100/ T-100 turbidimeter. The TN-100/ T-100 allow
Instruction Manual TN-100/ T-100 2Sample VialSample WellIndex MarkLiquid Crystal DisplayKeypad Figure 1: Parts of TN-100/ T-100 Turbidimeter 1.2
Instruction Manual TN-100/ T-100 31.3 Keys and Functions The keypad has five keys: ON/OFF, CAL, ▲, ▼, and READ/ENTER. Keys Functions ON/OFF -
Instruction Manual TN-100/ T-100 41.4 Battery Installation Four AAA-sized batteries are included in your meter’s packaging: 1. Use a Philips
Instruction Manual TN-100/ T-100 52 TURBIDITY CALIBRATION The TN-100/ T-100 turbidimeter was calibrated and tested prior to leaving the factory.