EUTECH INSTRUMENTS CYBERSCAN DO 5000 DO METER Instrukcja Obsługi Strona 54

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 76
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 53
Instruction Manual PD 300
8.12 P7.0 [6.0] rSt: Resetting to Factory Default Setting (DO)
This program lets you reset all DO (% saturation and mg/L) parameters to factory default
settings. This clears all calibration data and any other DO setup functions you might have
changed.
The following settings will remain as you have set them:
Temperature unit of measure (°C or °F)
The temperature offset calibration value
All DO calibration data and parameters
There are also some other parameters that retain settings when reset is done. See Addendum
3 on page 68 for a table of factory default settings.
From measurement mode
Figure 41: Reset to factory
defaults
1. Press SETUP key to enter SETUP mode.
2. Press or keys to scroll through su/jointfilesconvert/267593/bgroups
until you view parameter “rSt” in the upper display.
3. Press the ENTER key.
4. Press or keys to toggle between NO and
YES.
NO retains current memory.
YES clears all memory
5. Press the ENTER key to confirm selection and to
return to measurement mode.
49
Przeglądanie stron 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 75 76

Komentarze do niniejszej Instrukcji

Brak uwag