8.8 P3.0 [P2.0] (CAL): Previous Calibration Information
(*Program numbers in “[ ]” are for setup in mg/L (ppm) mode.
This sub group shows you the previous calibration data. This is a “view only” parameter.
In % saturation mode: Calibration information is shown in % saturation units.
In mg/L (ppm) mode: Calibration information is shown in mg/L (ppm) units.
From measurement mode:
1. Press the MODE key to select the calibration data
you want to view: % or mg/L (ppm).
2. Press the SETUP key to enter SETUP mode.
3. Press ▲ or ▼ keys to scroll through su/jointfilesconvert/267593/bgroups
until you view parameter “CAL” in the upper display.
4. Press the ENTER key. The upper display shows the
calibration data.
5. Press the ENTER key to move back to su/jointfilesconvert/267593/bgroup
“CAL”. If you want to return to measurement mode,
press CAL/MEAS.
NOTE: If you did not calibrate this meter in a particular
mode, the screen will show “ ---“. See Figure 30.
Figure 31: View calibration data
(The values shown above are for illustration purpose. Your meter may
display different values depending on the settings & your environmental
conditions)
Figure 32: eter is not calibrated