Instruction Manual DO 300
51
16 FACTORY DEFAULT SETTINGS
Resetting the meter to factory default settings clears all DO calibration data and memory, and
returns other setup functions to the default settings shown in the table below. However, the
clock settings will remain as you have set them. See page 35 for directions on setting the
meter to factory default settings.
Program Function Options/settings Default
CLr Memory clear Yes / no No
dPr Pressure inputs Hg or Pa Hg
Barometric pressure Adjust from 500 – 1499 mm
Hg
760 mm Hg
Salinity adjustment Adjust from 0 – 50.00 ppt 0 ppt
OFS Set % saturation offset Up to ±10% No offset
CAL View calibration data ---- ----
ELE View probe slope ---- 1.000
View % saturation offset ---- 0% offset
View mV = 100% saturation ---- 37.0 mV
View mV = 0% saturation ---- 0.3 mV
COF Ready indicator / auto endpoint On / off Ready ON only
Select mg/L or ppm units mg/L or ppm mg/L
Select temperature units °C or °F Retains settings
LCd Adjust back lit display Levels 0 to 8 (brightest) Retains settings
CLO Setting clock Year, date, time Retains settings
rSt Reset to factory default Yes / no no
NOTES:
The P7.0 clock function retains the year and time that you set, even after reverting to factory
default settings.