Instruction Manual CyberScan pH 1100/ 2100
49
10.4.8 Ion calibration data reset setup program P3.6
Use this program to erase all ion calibration data.
1. Enter ion setup mode as described in steps 1-4 in Section
10.4.1 Entering ion setup mode (page 43).
2. Use ∆/MI key to scroll to program Ion Cal Reset: P3.6. See
Figure 155.
3. Press ENTER to enter program Ion Cal Reset: P3.6.
4. Press >/CAL key to toggle between Ion Reset: Yes and
No. The selection will flash. See Figure 156.
5. Press ENTER to confirm selection.
• If you selected Yes, the screen flashes Ion being
Reset, Please wait... See Figure 157. The meter then
returns to setup mode.
• If you select No, the screen flashes Reset Aborted and
then returns to the setup mode. See Figure 158.
This is the last ion setup program. Press ∇/MR to go to the
previous program. Press ∆/MI to go back to the first ion
setup program. Press EXIT to return to mV measurement
mode.
Mode
Ion being Reset
Please wait. . .
Ion Reset: Yes No
Select & Press ENTER
Ion Cal Reset :P3.6
Press ENT. to Select
Figure 155
Figure 156
Figure 157
Figure 158
Reset Aborted