Instruction Manual CyberScan pH 1100/ 2100
39
10.3.3 mV calibration due alarm setup program P2.1
Use this program to set a reminder when the next calibration is
due. Choose from 1 to 365 days from last calibration.
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use /MI key to scroll to program mVCalDue Alarm: P2.1.
See Figure 116.
3. Press ENTER to enter program mVCalDue Alarm: P2.1.
4. Press >/CAL to toggle between Cal Alarm Due:On and Off.
The selection will flash. See Figure 117.
5. Press ENTER to confirm selection.
If you selected On, continue to step 6.
If you select Off, pressing ENTER returns to the setup
mode.
Press /MI to continue to the next setup program, or press
/MR to go to the previous program. Press EXIT to return to
mV measurement mode.
6. If you selected On in step 5, the display reads Key in Cal
Due Period (1-365 days) =. Key number of days using the
numeric keypad. If you make a mistake, use the /MI or
/MR keys to highlight mistake, then re-enter.
See Figure 118.
7. Press ENTER to confirm number of days to next calibration,
screen flashes Sys Updating, Please wait... and returns to
the setup mode.
Press /MI to continue to the next setup program, or press
/MR to go to the previous program. Press EXIT to return to
mV measurement mode.
Keyin Cal Due Period
(1-365 days) = 1
Cal Alarm Due On Off
Select & Press ENTER
mVCalDue Alarm :P2.1
Press ENT. to Select
Figure 116
Figure 117
Figure 118