
Instruction Manual CyberScan pH 1100/ 2100
45
10.4.4 Ion calibration due alarm setup program P3.2
Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.
1. Enter ion setup mode as described in steps 1-4 in Section 10.4.1
Entering ion setup mode (page 43).
2. Use ∆/MI key to scroll to program IonCalDue Alarm: P3.2.
See
Figure 138.
3. Press ENTER to enter program IonCalDue Alarm: P3.2.
4. Press >/CAL key to toggle between Cal Alarm Due: On and Off.
The selection will flash. See Figure 139.
5. Press ENTER to confirm selection.
• If you selected On, continue to step 6.
• If you select Off, pressing ENTER returns to the setup mode.
Press ∆/MI key to continue to the next setup program, or press
∇/MR to go to the previous program. Press EXIT to return to mV
measurement mode.
6. If you selected On in step 5, the display reads Key in Cal Due
Period (1-365 days) = 1. Key number of days using the numeric
keypad. If you make a mistake, use ∆/MI or ∇/MR keys to
highlight mistake, then re-enter. See Figure 140.
7. Press ENTER to confirm number of days to next calibration;
screen flashes Sys Updating, Please wait… and returns to the
setup mode. Press ∆/MI key to continue to the next setup
program, or press ∇/MR to go to the previous program. Press
EXIT to return to mV measurement mode.
Keyin Cal Due Period
(1-365 days) = 1
Cal Alarm Due: On Off
Select & Press ENTER
IonCalDue Alarm :P3.2
Press ENT. to Select
Figure 138
Figure 139
Figure 140
Komentarze do niniejszej Instrukcji