13 ADDENDUM 1: METER FACTORY DEFAULT
SETTINGS
Resetting the meter to factory default settings clears all calibration data and most other setup
functions you might have changed. The following settings will remain as you have set them:
Temperature unit of measure (°C or °F) – only for pH 310 meter
The temperature offset calibration value.
Clock function only for pH 310 meter
Type Parameter P No. Default Remarks
CLr Memory clear P1.0 No Retains your current settings
CAL View pH calibration data P2.1 –– Calibration data for 1st buffer
P2.2 –– Calibration data for 2nd buffer
P2.3 –– Calibration data for 3rd buffer
P2.4 –– Calibration data for 4th buffer
P2.5 –– Calibration data for 5th buffer
(only in pH 310 meter) P2.6 –– Calibration data for 6th buffer (DIN only)
ELE View electrode offset P3.1 0.00 mV No offset adjustment
View electrode slope P3.2 100.0 % No slope adjustment
View relative mV offset P3.3 0 mV No mV adjustment
COF Ready indicator P4.1 Ready On Ready indicator on; auto endpoint off
# pH calibration points P4.2 3 3 pH calibration points available (1 – 5 pt
range)
Calibration buffer selection
sets(pH 310 meter only)
P4.3 USA
°C or °F (pH 310 meter) P4.4 No default °C or °F remains as selected
CLO Setting year, date and time
(in pH 310 meter only)
P5.0 No default Year, date and time remain as selected
rSt pH factory default
For pH 300 meter
For pH 310 meter
P5.0
P6.0
No
Retains your current settings