EUTECH INSTRUMENTS CON 700 - REV 2 Instrukcja Obsługi Strona 27

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 26
22
14. SPECIFICATIONS
pH Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy
0.01 pH + 1 count
Calibration Points Up to 5 points with Auto-buffer recognition
Buffer Options
USA : pH 1.68, 4.01, 7.00, 10.01,12.45
NIST: pH 1.68, 4.01, 6.86, 9.18 ,12.45
Slope Display Yes (with offset)
mV Range
2000 mV
Resolution
0.1 mV ( 199.9 mV), 1 mV beyond 200 mV
Accuracy
0.2 mV ( 199.9 mV), 2 mV beyond 200 mV
Offset Adjustment Up to ±150 mV
Conductivity Range
0 to 20.00, 200.0, 2000 µS
0 to 20.00, 200.0 mS
TDS Range
(0.5 TDS factor)
0 to 10.00, 100.0, 1000 ppm
0 to 10.00, 100.0 ppt
(max 200.0 ppt @ 1.0 TDS factor)
Resolution 0.05 % Full Scale
Accuracy ±1% Full Scale
Temperature
Range (Meter) 0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Compensation
Type & Range
Automatic with supplied cell or Manual.
0.0 to 100 ºC / 32.0 to 212.0 ºF,
0.0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell
Resolution 0.1 ºC / ºF
Accuracy ± 0.5ºC / ± 0.9ºF
Calibration
Offset in 0.1 º increments; Offset range: 5 ºC / 9 ºF
Coefficient 0.0 to 10.0% per ºC
Normalization 15.0 to 30.0 ºC (adjustable)