Press .
Press or to select the desired value (15.0 to 30
Press
.0).
to confirm.
ACAL to nductivity Calibration)—Con & TDS
only9.11
3.7 (Au Co
See Sectio for more deta atic Conductivity Calibration.
ress
n 5.1 ils on Autom
. P
Press or to select “Yes” (Automatic) or “No” (Manual).
Press to confirm.
9.12 3.8 Po
See Sectio for details o
n Single Point & Multi Point Calibration.
ress
SPC (Single int Calibration)—Con & TDS only
n 5.2
. P
Press or to select “Yes” (Single) or “No” (Multi).
Press to confirm.
9.13 3.8 No al
The PC 700 includes a probe with a nominal cell constant (k) of 1.0.
Use prob k parately) for improved
performa ce in extreme sample ranges. Use this setup function to
atch the
CELL ( min Cell Constant)—Con & TDS only
es with = 0.1 and 10 (sold se
n
change the cell constant if necessary. Meter default is 1.0 to m
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).
Press
.
Press or to select 0.1, 1.0, or 10.0
Press to confirm.