© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
If Auto-range was selected then Aout max1 can be configured. Aout max1 is the maximum
value for the first range on Auto-Range. The maximum value for the second range on Auto-
Range was set in the previous menu. If Logarithmic Range was selected, it will also prompt for
the number of decades as ‘Aout1 # of Decades =2’.
The value for the Hold mode can be configured to hold the last value or can be set to a Fixed
value.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will dis-
card the entered values and return to the measurement display screen, selecting Yes will save
changes made.