1 INTRODUCTION
1.1 Description of Unit
Thank you for purchasing Eutech’s ¼ DIN alpha-800 series pH/ORP
process controllers. This unit is used for measuring either pH or ORP
parameter one at a time, and the operational mode is switchable from the
menu. You can use this unit to measure pH or ORP with limit control.
This controller has many user-friendly and safety features which include:
Menu-driven program that simplifies set-up
Built-in non-volatile memory backup to ensure that calibration and
other information are not erased if power supply fails
Push-button two-point calibration and electrode offset
adjustment from the keypad
Automatic temperature compensation (ATC)
Manual temperature compensation setting without the ATC probe,
with independent setting for calibration and process temperature
0 to 2000 second time delay adjustment on all relays – minimise
false alarms
Separately adjustable high and low set point hysteresis (dead
bands) prevent chattering of relays around the set points
Asymmetrical/symmetrical input for pH/ORP operation
Large dual display LCD for easy reading with clear multiple
annunciators, alarm status, operational and error messages
Two switching contacts as set-point triggering relays
Hold function freezes output current (4/20mA) and control relays
LED indicators signal control activities to monitor controller status
from a distance
Protection against electromagnetic interference – Input and
output are free from external inference