7.4 Configuration (ConF) sub-function
The options available in this sub-function allows the controller to be
configured as a pH or ORP controller.
7.4.1 Entering the Configuration sub-function
1) Enter Advanced Set-up mode. Press the ENTER key.
2) Press the ∆
∆∆
∆ or ∇
∇∇
∇ keys to scroll until the upper display shows “ConF”.
7.4.2 Selecting pH or ORP mV measurement
Besides selecting the controller to function as a pH or ORP (mV) controller, it
also allows the selection of the type of input: symmetrical or asymmetrical.
Symmetrical mode is used in conjunction with a potential matching pin – in
an electrically noisy environment. Please refer to section 3, page 6 for a
graphical representation.
1) Follow directions in 7.4.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the ∆
∆∆
∆ or ∇
∇∇
∇ keys until the upper display
shows the mode of operation (pH or ORP mV and the lower display shows
symmetrical (SY) or asymmetrical (ASY) input type.
3) Proceed to 7.4.3, or return to Measurement mode by pressing the ∆
∆∆
∆ and
∇
∇∇
∇ keys simultaneously (escape).
SETUP
ConF
HOLD
PH
SY
SETUP
HOLD
PH
ASY
SETUP
HOLD
SETUP
HOLD
OrP
mV
SY
SETUP
HOLD
OrP
mV
ASY