5 MEASUREMENT
5.1 Display in Measurement mode
When the controller is initially powered on, it automatically enters into the
Measurement mode after the large dual LCD displays all segments briefly.
The upper display shows the measured pH or ORP value, while the lower
display shows either the temperature value if the controller is set for pH
measurement or “OrP” if it is set for ORP measurement.
Annunciators at the right side of the display indicate whether the controller is
set for pH or mV measurement..
Similarly annunciators or icons at the top or left side of the display shows the
current status of controller, e.g. “MEAS”, “SETUP”, “CAL”, “READY”, etc. In
addition, error messages also guide the user in time of any faulty conditions
as indicated by “ERR” or icons flashing.
5.1.1 Check electrode performance
To read current electrode slope and offset values:
1) Press the CAL key followed by the ENTER key. The upper display
shows electrode slope. The lower display reading shows the pH reading at 0
mV.
2) Press the
and
keys to return to Measurement mode.