Instruction Manual CyberScan PC5500/ 5000
132
clear
Standardize - Ion
setup
Touch to measure sample
or
meas
Touch to access standardize mode
std
std
meas
mV
STABLE
STND
Not Standardized
ISA Volume
1 2 3
4 5 6
7 8 9
- 0 .
BS exit enter
channel 1
clear
Standardize - Ion
setup
Touch to measure sample
or
meas
Touch to access standardize mode
std
std
meas
mV
STABLE
STND
Not Standardized
Concentration
1 2 3
4 5 6
7 8 9
- 0 .
BS exit enter
channel 1
Calculated Slope:
Reading 1:
Reading 2:
Sample Volume:
Standard Volume:
ISA Volume:
Standard Concentration:
Known Addition Result
OK print
58 mV
222.7 mV
200.6 mV
50
10
50
100 ppm
12.2
ppm