Instruction Manual
CON 510
Bench Conductivity/TDS Meter
Tech nol o
g
y
M
ade
E
a
s
y
.
.
.
68X090820
Rev. 0 11/02
Przeglądanie stron 0
1 2 3 4 5 6 ... 33 34

Podsumowanie treści

Strona 1 - CON 510

Instruction Manual CON 510 Bench Conductivity/TDS Meter Tech nol ogyMadeEasy...68X090820 Rev. 0 11

Strona 2 - Preface

Instruction Manual CON 510 7 4.6 Automatic Calibration (Only for Conductivity) In the automatic calibration mode, the CON 510 bench meter is capab

Strona 3

Instruction Manual CON 510 8 4.7 Manual Calibration (Conductivity/TDS) The meter accepts either single point or multi-point (up to 5 points) man

Strona 4 - -8.8.8.8

Instruction Manual CON 510 9 4.8 Temperature Calibration The Conductivity electrode (EC-CONSEN91W/ 35608-50) has a built-in temperature sensor f

Strona 5 - 2.2 Keypad

Instruction Manual CON 510 10 5 MEASUREMENT The CON 510 bench meter measures up to 5 different measurement ranges with auto-ranging capability whi

Strona 6 - 3 PREPARATION

Instruction Manual CON 510 11 5.2 Manual Temperature Compensation NOTE: For manual temperature compensation, you must deactivate the ATC mode to

Strona 7

Instruction Manual CON 510 12 5.4 Using Manual Ranging Function Although the CON 510 bench meter has an automatic ranging capability, you may al

Strona 8 - 4 CALIBRATION

Instruction Manual CON 510 13 6 MEMORY AND DATA INPUT FUNCTIONS 6.1 Memory Input The meter’s non-volatile memory can store and recall up to 50 da

Strona 9

Instruction Manual CON 510 14 7 SETUP FUNCTIONS The setup mode allows you to customize the meter’s setting to your individual preferences. The CON

Strona 10

Instruction Manual CON 510 15 SETUP . P5.0: Mode of Calibration Selection of Automatic or Manual Calibration (only in Conductivity mode) Sele

Strona 11

Instruction Manual CON 510 16 7.3 P2.0: Viewing Electrode Diagnosis Main program 2.0 shows the effective cell constant for each range being calib

Strona 12

Preface This manual serves to explain the use of the CON 510 bench meter. It functions in two ways, firstly as a step by step guide to help you to o

Strona 13 - 5 MEASUREMENT

Instruction Manual CON 510 17 P3.2 Selecting °C or °F You can select between °C and °F as unit of measure for temperature readings. Meter default

Strona 14 - 5.3 Taking Measurements

Instruction Manual CON 510 18 P3.4 Setting the TDS factor (In TDS Mode) The concentration of salts dissolved in solution increases the conducti

Strona 15 - 5.5 HOLD Function

Instruction Manual CON 510 19 P4.2 Adjusting the Normalization Temperature Your meter will normalize its conductivity measurements to a standard

Strona 16 - 6.2 Memory Recall

Instruction Manual CON 510 20 7.6 P5.0 Mode of calibration The CON 510 bench meter allows you to perform automatic (only for Conductivity mode) o

Strona 17 - 7 SETUP FUNCTIONS

Instruction Manual CON 510 21 7.7 P6.0 Selecting the cell constant This program allows you to select the appropriate cell constant of K = 0.1, 1.

Strona 18

Instruction Manual CON 510 22 7.8 P7.0: Resetting to factory default settings Program 7.0 allows you to reset all parameters to factory default s

Strona 19

Instruction Manual CON 510 23 8 PROBE CARE AND MAINTENANCE Keep the Conductivity/TDS probe clean. Rinse the probe twice, and gently swirl it whil

Strona 20

Instruction Manual CON 510 24 9 TROUBLE SHOOTING GUIDE Problem Possible Cause Solution No display A. AC outlet not switched on. A. Switch on

Strona 21 - 7.5 P4.0: Temperature

Instruction Manual CON 510 25 11 SPECIFICATIONS SPECIFICATIONS CON510 Conductivity Range 0 to 20.00, 200.0, 2000 µS/cm; 0 to 20.00, 200.0 mS/cm T

Strona 22

Instruction Manual CON 510 26 12 ACCESSORIES Replacement Meter and Accessories Item Description Eutech Instruments Order Code No. Oakton Inst

Strona 23

TABLE OF CONTENTS 1 INTRODUCTION 1 2 DISPLAY AND KEYPAD FUNCTIONS 1 2.1 Display 1 2.2 Keypad 2 3 PREPARATION 3 3.1 Conductivity Electrode

Strona 24

Instruction Manual CON 510 27 13 NOTE: CONDUCTIVITY AND TDS SOLUTIONS HAVE ±1% ACCURACY AT 25°C ADDENDUM 1: CALIBRATION TIPS You only need one

Strona 25

Instruction Manual CON 510 28 15 ADDENDUM 3: CALCULATING TEMPERATURE COEFFICIENTS To determine the temperature coefficient of your sample soluti

Strona 26

Instruction Manual CON 510 29 16 ADDENDUM 4: METER FACTORY DEFAULT SETTINGS Type Parameter Default Remarks P1.1 Viewing calibration solution

Strona 27 - 10 ERROR MESSAGES

Instruction Manual CON 510 30 17 WARRANTY The CON 510 bench meter is supplied with a 3-year warranty from manufacturing defects and electrodes for

Strona 28 - 11 SPECIFICATIONS

31 For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distribut

Strona 29 - 12 ACCESSORIES

Instruction Manual CON 510 1 1 INTRODUCTION Thank you for selecting the CON510 bench meter. This meter is a microprocessor-based instrument that i

Strona 30

Instruction Manual CON 510 2 2.2 Keypad The splash-proof membrane tactile keypad allows easy key entry. Each button, when pressed, has a correspon

Strona 31

Instruction Manual CON 510 3 3 PREPARATION 3.1 Conductivity Electrode Information The CON510 bench meter is supplied with a Conductivity/TDS elec

Strona 32

Instruction Manual CON 510 4 3.4 Connecting the Electrode Holder The integral electrode holder serves as a handy holder for a

Strona 33 - 18 RETURN OF ITEMS

Instruction Manual CON 510 5 4 CALIBRATION 4.1 Important Information on Meter Calibration The CON 510 bench meter allows you to perform automat

Strona 34

Instruction Manual CON 510 6 Normalization Temperature: The factory default value for normalization temperature is 25 °C. If you need to normalize