32 Thermo Scientific Alpha pH 560
11
From pH measurement mode press ENT key to enter setup
mode as described in page 24. The LCD shows the first screen
of setup mode (OFS). Press or key to select buffer
settings screen (bUFF).
Press ENT key to access buffer settings (bUFF).
2
Selecting buffer group: The lower display shows the last
configured buffer group. Press or key to select the
required buffer group (‘USA’ or ‘nST’).
Press ENT key to confirm your selection. The pH Controller
reverts to bUFF screen.
Press or key to access other setup screens or Press and
key simultaneously (escape) to return to measurement mode.
Configuration settings let you configure the pH Controller to different
measurement & operation modes, enable/disable backlight of the LCD
& reset the pH Controller to factory defaults.
Asymmetrical Operation Mode (ASY):
This is the default mode of operation of the pH Controller, which is
suitable for most applications where sample (or process) liquid is not
interfered by adjacent electrical noise.
Symmetrical Operation Mode (SY):
This is mode of operation is recommended for applications where
sample (or process) liquid is subjected to high electrical interference and
therefore the pH reading becomes unstable. When the pH Controller is
set to symmetrical mode, you should immerse the potential matching
pin (PMP) in the sample (or process) liquid or use a pH electrode with
built-in potential matching pin. The ‘MEAS’ indicator blinks when the
pH Controller is set to Symmetric mode.
Configura
t
ion
Settings