Alpha pH 560 pH Controller pH/mV/ºC meter

Strona 2 - Preface

Thermo Scientific Alpha pH 560 5 Figure 2–2. Outer Side of Back PCBA The Controller supports both Pt100 & Pt1000 (2-wire or 3-wire) tempera

Strona 3 - Contents

6 Thermo Scientific Alpha pH 560 Figure 2–3. Inner Side of Back PCBA Set the J7 jumper to required sensor (Pt100 or Pt1000) type. Set Jump

Strona 4

Thermo Scientific Alpha pH 560 7 1. If the pH/ORP electrode has a BNC connector, remove the BNC connector from the cable. NOTE: Oakton Instrum

Strona 5

8 Thermo Scientific Alpha pH 560 Figure 2–5. Frontal View Figure 2–6. Side View Installation Mechanical Dimensions

Strona 6 - Introduction

Thermo Scientific Alpha pH 560 9 Figure 2–7. Figure 2–8. Wall Mount Panel Mount

Strona 7 - Correct Disposal

10 Thermo Scientific Alpha pH 560 Figure 2–9.

Strona 8 - Getting Started

Thermo Scientific Alpha pH 560 11 The Liquid Crystal Display (LCD) of Alpha pH 560 Controller has two alpha-numerical displays (Upper and a Lowe

Strona 9

12 Thermo Scientific Alpha pH 560 Status Annunciators READY Appears when the reading is stable HOLD Appears in Setup mode and Calibration mod

Strona 10

Thermo Scientific Alpha pH 560 13 Figure 2–11. The Liquid Crystal Display (LCD) of Alpha pH 560 Controller has two alpha-numerical displays

Strona 11

14 Thermo Scientific Alpha pH 560 Chapter 3 Operation When the pH Controller is powered on, the display shows all the LCD segments briefly, and

Strona 12 - Terminals

Thermo Scientific Alpha pH 560 Preface This manual serves to explain the use of the Thermo Scientific Alpha pH 560 Controller. The manual functio

Strona 13 - Installation

Thermo Scientific Alpha pH 560 15 Figure 3–13. Menu Overview

Strona 14 - Panel Mount

Thermo Scientific Alpha pH 560 16 Chapter 4 Calibration Mode Before starting calibration, make sure that the pH Controller is in appropriate me

Strona 15

Thermo Scientific Alpha pH 560 17 NOTES: • To exit calibration mode at any time during calibration, Press ▲ and ▼ keys simultaneously (escape)

Strona 16 - Display &

18 Thermo Scientific Alpha pH 560 2 Place the electrode in pH 7.00 buffer. Immerse the temperature probe in the buffer solution if ATC mode is e

Strona 17 - Status Annunciators

Thermo Scientific Alpha pH 560 19 2 Place the electrode in pH 7.00 buffer. Immerse the temperature probe in the buffer solution if ATC mode is e

Strona 18 - Key Functions

20 Thermo Scientific Alpha pH 560 calibration mode, the pH Controller reverts back to ‘process temperature’. Make sure that the pH Controller is

Strona 19 - Operation

Thermo Scientific Alpha pH 560 21 Figure 4–18. NIST Buffer 2-Point Calibration 11 From pH measurement mode press CAL key to enter calibration

Strona 20 - Menu Overview

22 Thermo Scientific Alpha pH 560 Press ENT key to exit from the calibration. 55 The pH Controller reverts to pH measurement mode. NOTES: • To

Strona 21 - Calibration Mode

Thermo Scientific Alpha pH 560 23 ORP value of the solution. Allowable range: -150mV to 150mV. 3 Press ENT key to confirm the value. 4 The pH Co

Strona 22 - For USA Standard

Thermo Scientific Alpha pH 560 24 Chapter 5 Setup Mode The setup mode allows you to customize the settings of the pH Controller to suite your r

Strona 23

Thermo Scientific Alpha pH 560 Contents Introduction ...

Strona 24

Thermo Scientific Alpha pH 560 25 NOTE:‘Electrode Offset Setting’ is not available when the pH Controller is configured for ORP measurement mode

Strona 25 - For NIST Buffer

26 Thermo Scientific Alpha pH 560 Controller adjust the lower reading automatically to suit the new offset value. Up to ± 2.00 pH offset is allo

Strona 26

Thermo Scientific Alpha pH 560 27 Press ▲ or ▼ key to select the desired units for temperature (ºC or ºF). Press ENT key to confirm your selecti

Strona 27 - ORP Calibration

28 Thermo Scientific Alpha pH 560 to adjust the upper display to desired calibration temperature. Allowable range: –10.0 to 125.0°C / 14.0 to 25

Strona 28 - Calibration

Thermo Scientific Alpha pH 560 29 configured set point value, otherwise the default value. Press ▲ or ▼ key to enter the value for set point 1

Strona 29 - Setup Mode

30 Thermo Scientific Alpha pH 560 Press ▲ or ▼ key to access other setup screens or Press ▲ and ▼ key simultaneously (escape) to return to meas

Strona 30 - Settings

Thermo Scientific Alpha pH 560 31 parameters. Press ENT key to confirm your selection. If Controller type was set to Off (OFF) The pH Co

Strona 31

32 Thermo Scientific Alpha pH 560 11 From pH measurement mode press ENT key to enter setup mode as described in page 24. The LCD shows the first

Strona 32 - ATC’ annunciator

Thermo Scientific Alpha pH 560 33 Figure 5–26. Configuration Settings 11From pH (or ORP) measurement mode press ENT key to enter setup mode as

Strona 33 - Setting

34 Thermo Scientific Alpha pH 560 to their default values. 4 Select operation mode: The upper display shows the last configured operation mode.

Strona 34

Thermo Fisher Scientific iii For NIST Buffer ... 20 1-Point Calibration:

Strona 35 - Controller

Thermo Scientific Alpha pH 560 35 Each time you calibrate your pH electrode, the pH Controller re-calculates slope & offset of the electrode

Strona 36 - Buffer Selection

Thermo Scientific Alpha pH 560 36 Chapter 6 Technical Specifications Product Technical Specification GENERAL SPECIFICATIONS (a) pH Measuring Ra

Strona 37 - Configura

Thermo Scientific Alpha pH 560 37 Product Technical Specification Number of calibration points 1 or 2 points Number of calibration buffers USA:

Strona 38

Thermo Scientific Alpha pH 560 38 Chapter 7 List of Accessories Controller Replacement and Accessories Item Description Order Code Alpha-pH 56

Strona 39

Thermo Scientific Alpha pH 560 39 NOTE: • pH buffer solutions (480 ml bottle) have ±0.01 pH accuracy at 25 °C • pH buffer sachets are individu

Strona 40 - Properties

Thermo Scientific Alpha pH 560 40 Chapter 8 Troubleshooting Problem Cause Solution Power on, but no display a) Loose connections b) Incorrect

Strona 41 - Technical Specifications

Thermo Scientific Alpha pH 560 41 Chapter 9 General Information Thermo Fisher Scientific warrants this product to be free from significant devi

Strona 42 - ± 10 ˚C/ ± 18 ˚F

42 Thermo Scientific Alpha pH 560 Appendices The following table shows the various pH values at different temperature of the solution during cal

Strona 43 - List of Accessories

Index: Error! No text of specified style in document. Thermo Scientific Alpha pH 560 43 The relay of the pH Controller activates when the set-po

Strona 44 - Instruments

44 Thermo Scientific Alpha pH 560 Abbreviation Description ASy Asymmetrical mode AtC Automatic Temperature Compensation bUFF Buffer CAL Cali

Strona 46 - General Information

Index: Error! No text of specified style in document. Thermo Scientific Alpha pH 560 45 Abbreviation Description Set Setting SP1 1st Set point S

Strona 47 - Appendices

46 Thermo Scientific Alpha pH 560

Strona 48 - Appendix 3

Water Analysis InstrumentsNorth America166 Cummings CenterBeverly, MA 01915 USAToll Free: 1-800-225-1480Tel: 1-978-232-6000Dom. Fax: 1-978-232-6015Int

Strona 49 - Appendix 4

Thermo Scientific Alpha pH 560 1 Chapter 1 Introduction Thank you for purchasing the Alpha pH 560 Controller. The construction of the pH 560

Strona 50

2 Thermo Scientific Alpha pH 560 • The Alpha pH 560 must only be operated under the specified operating conditions (see Page 36). • The Alpha

Strona 51

Thermo Scientific Alpha pH 560 3 Chapter 2 Getting Started The Thermo Scientific Alpha pH 560 Controller is used for measuring pH and temperatur

Strona 52 - Water Analysis Instruments

4 Thermo Scientific Alpha pH 560 A typical measurement system consists of: • A pH process Controller • A pH sensor with integrated or separa