5.7 Configuration Settings
Configuration settings let you to enable/disable backlight of the LCD, select
between 2-cell & 4-cell and reset the controller to factory defaults.
ENT
SETUP
HOLD
2
SETUP
HOLD
2
ENT
ENT
SETUP
HOLD
3
SETUP
HOLD
3
ENT
SETUP
HOLD
SETUP
HOLD
5
4
ENT
ENT
SETUP
HOLD
ENT
ENT
SETUP
HOLD
5
SETUP
HOLD
5
ENT
1
MEAS
°C
ATC
READY
µS
This screen appears only
when 2-cell is selected
and measuring range is set to
0.0 to 200.0 mS/cm
R.
11
From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press or key to select configuration settings screen
(Cnfg).
Press ENT key to access configuration settings (Cnfg).
2
Enable/disable LCD backlight: The upper display shows ‘LitE’. The
lower display shows last configured backlight selection (‘On’ or ‘OFF’).
Press or key to enable (LitE On) or disable (LitE OFF) backlight
Przeglądanie stron 32