24
5.5 Controller Settings
The controller settings allow you to set the relay type and relay status of the
controller.
ENT
SETUP
HOLD
2
ENT
SETUP
HOLD
3
SETUP
HOLD
1
SETUP
HOLD
2
ENT
SETUP
HOLD
3
ENT
ENT
11
From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup
mode (rAng). Press ▲ or ▼ key to select controller settings screen
(Cntr).
Press ENT key to access controller settings (Cntr).
2
Select the controller type: The lower display shows ‘tyP’. The upper
display shows the last configured controller type. (‘OFF’ or ‘L.Ct’).
Press ▲ or ▼ key to select the desired controller type.
OFF :
Controller Off
Selecting ‘OFF’, the conductivity controller stops
functioning as a controller by preventing relay functions.
Once the controller type is set to ‘OFF’, the setup mode
does not show you the Relay 1/Relay 2 settings.
L.Ct: Limit value control (Controller On)
Selecting ‘L.Ct’, the conductivity controller is set to function
as a controller. When the controller type is set to ‘L.Ct’, the
Relay 1/Relay 2 settings are available in the setup mode
for you to configure relay operating parameters.
Press ENT key to confirm your selection.