Seven ranges in a single controller allows the
controller to be used in a wide variety of
applications. Maximum measurement range is
1000 mS/cm. Accuracy of ±1% is achievable
for most ranges, while ±2% for measurements
above 500 mS/cm.
4- or 2-cell electrode system can be used with a
input of cell constants, from 0.005 to 9.999.
New 4-cell electrode system eliminates polarizing
effects ensuring more accurate measurements in
highly conductive samples.
The cell constant corresponding to the
cell connected to the controller can be input
independently during calibration.
Status of the cell is updated and displayed after
every calibration and can be viewed at anytime.
Temperature Coefficients can be set for more
precise temperature compensation. Select from
0.0% to 10.0%.
In addition, for 0 to 2.000 µS/cm range, select ‘Pure
water’ compensation graph to correct for the non-
linearity of pure water temperature correction
curves.