2 PRODUCT DESCRIPTION
2.1 Description of unit
The αlpha pH 2000D is used for measuring pH and temperature values. The pH values can be
measured using industrial differential & combination pH sensors. The temperature values can
be measured using 2-wire / 3-wire Pt100 / Pt1000 sensors or NTC300 thermistor. The
transmitter is available in a wall mounting enclosure. The controller / transmitter can be used
for applications such as water treatment and monitoring, galvanic-decontamination, chemical
processing, food processing, clean or wastewater control and neutralization processes.
This transmitter has many user-friendly and safety features which include:
Push-button for calibration and sensor offset adjustment from the keypad.
Built-in non-volatile memory to ensure that calibration and other information are not
erased if power supply fails.
Menu-driven program that simplifies set-up.
Automatic temperature compensation (ATC).
Manual temperature compensation setting without the ATC probe, with independent
setting for calibration and process temperature.
Two galvanically isolated current outputs 0/4...20mA.
Two switching contacts as set-point relays.
0 to 2000 seconds time delay adjustment on all relays – minimizes false alarms.
Separately adjustable high and low set-point hysteresis (dead bands) prevent chattering
of relays around the set points.
Three control modes: limit controller, P controller and PI controller (P/PI controller as
pulse length or pulse frequency).
Large dual display LCD for easy reading with clear multiple annunciators, alarm status,
operational and error messages.
Separate alarm relay alerting you to set point limits exceeded for a certain time and if the
Pt100/Pt1000 or NTC300 thermistor wires are broken or disconnected during the ATC
function.
Wash relay.
Hold function to freeze output current (0/4...20mA) and release control relays.
LED indicators signal control activities to visually monitor transmitter status from a
distance.
Protection against electromagnetic interference.
Back lit and UV light protected LC display.