Instruction Manual Salt 6
25
7 PROBE CARE AND MAINTENANCE
Keep the conductivity probe clean. Rinse the probe twice, and gently swirl it
while you take readings. For best accuracy, soak a dry probe for at least 5 to 10
minutes or longer before calibration. Rinse the probe with deionized water
before storing. Never scratch the bands with a hard substance. Do not strike the
probe against any hard surface.
Do not make continuous contact with your
solutions. Readings will rise over a period of time
while you soak your probe.
Do not immerse the probe in oily solutions. Clean
the electrode thoroughly by stirring it in a mild
detergent bath or isopropyl alcohol. Wipe the
probe with soft tissue paper. Rinse thoroughly in
tap water and then in deionized water. Recalibrate
the meter after cleaning the probe.
The conductivity probe (Order Part No. EC-
CONSEN91B/ 35606-55) which is included with
your meter features a removable probe guard to
make cleaning easy.
To remove probe guard:
1) Grip yellow probe guard and twist
clockwise. The locking notch will
release.
2) Slide probe guard off end of probe.
NOTE: Remember to re-attach the probe guard prior to taking readings. Failure
to do so could result in erroneous readings.