Bench700Series
Specication Tables
<<Wastewater&WaterTreatment
>>EcologicalStudies
pH/Ion
BenchMeters
pH700 Ion700
pH
Range -1.99to16.00pH 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy ±0.002
Cal. Points Upto5points
Buer Sets USA,NIST
ORP
Range ±1999 mV
Rel. mV Range ±1999 mV
Resolution 0.1 mV (±199.9 mV) / 1 mV (beyond)
Accuracy ±0.2 mV (±199.9 mV) / ±2 mV (beyond)
Ion
Concentration – 0.01 to 2000 ppm
Resolution (ppm) – 0.01 (<1); 0.1 (1.0 to 199.9); 1 (>200)
Accuracy – ±0.5%fullscale(monovalent)±1.0%fullscale(divalent)
Cal. Points – From 2 to 5 consecutive points; 0.1, 1.0, 10.0, 100.0 or 1000.0 ppm
Temperature
Range 0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Resolution 0.1 ºC / 0.1 ºF
Accuracy ±0.5 ºC / ±0.9 ºF
Compensation Automatic or Manual (0 to 100 ºC / 32 to 212 ºF)
Calibration Osetin0.1ºincrements;osetrange:±5ºC/9ºF
Conductivity
BenchMeter
CON700
Conductivity
Range .. µS/cm to 200.0 mS/cm
Resolution 0.01 μS; 0.1 μS ; 1 μS ; 0.01 mS ; 0.1 mS
Accuracy ±1%fullscale
Cal. Points Automatic (4 points); maximum 1 per range; manual (5 points); maximum 1 per range
Cell Constant 0.1, 1.0, 10.0 (selectable)
TDS
Range ..to100ppt@0.5TDSfactor(to200ppt@1.0TDSfactor)
Resolution 0.01 ppm; 0.1 ppm ; 1 ppm ; 0.01 ppt ; 0.1 ppt
Accuracy ±1%fullscale
Cal. Points Upto5points
TDS Factor 0.40 to 1.00 (adjustable)
Temperature
Range (Meter) 0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Resolution 0.1
º
C
Accuracy ± 0.5
º
C / ± 0.9
º
F
Coecient (Per ºC) 0.00to10.00%
Normalization 15.0 to 30.0 ºC (adjustable)
Compensation
Type & Range
Automatic with supplied cell or
manual 0.0 to 100 ºC / 32.0 to 212.0 ºF, 0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell
Calibration Osetin0.1ºincrements;osetrange:±5ºC/9ºF