6.3 ORP – % Calibration
The ORP % mode allows you to calibrate at two points: a low concentration sample (20%) and a
high concentration sample (80%).
To calibrate the controller for ORP %:
1) Enter Calibration mode. While in the Measurement
mode, press the CAL key and scroll to Calibration
code “11”. Press the ENTER key again. The upper
and lower display reads “CAL OrP”.
NOTE: If the display reads “CAL pH”, see section 7.6.2
for procedures on how to switch from pH to ORP mV
readings.
2) Press the ENTER key to begin calibration. The
“CAL” indicator appears on the display. The upper
display shows “80”. The lower display shows “C1
(=calibrated value 80%).
3) Place sensor in the high concentration sample
(relative value of 80%).
4) Press the ENTER key to confirm the value. The
electrode indicator and CAL indicator flash. The unit
automatically adjusts its reading to match the
solution value.
5) The display will now show ‘20’ and ‘C2’. Remove the
electrode from the first standard, wash it well and
then immerse it in the 20% standard. When the
display stabilises, press the ENTER key. The ‘CAL’
indicator flashes and the controller adjusts its display to the 20% relative value and exits to
Measurement mode.
Note: If there is a calibration error, the controller displays “ERR”, and it will not automatically return
to Measurement mode. If this happens, press the and keys simultaneously (escape) to
restart the calibration beginning from step 1.
Note: If you stop calibration after entering the first value, or if calibration is defective, the controller
will use its original calibration data.
CAL
SETUP
HO LD
80
SETUP
HO LD
C.1
20
SETU P
HO LD
C.2