Instruction Manual alpha-pH 1000
24
7.5 Current Output (rng) sub-function
This sub-function lets you set the controller current output range of this
unit.
7.5.1 Entering current output sub-function
1) Enter Advanced Set-up mode. Push the ENTER key and scroll to
Advanced Set-up security code “22”. Push the ENTER key again.
2) Press the or key to scroll until the upper display shows “rng”.
7.5.2 Choosing the output type
This parameter lets you choose between 0-20 mA or 4-20 mA output.
1) Follow directions in 7.5.1 to enter Current Output mode.
2) Press the ENTER key. Scroll with the or key until the upper
display shows the output type (0-20 or 4-20), and the lower display
shows “out”.
3) Press the or key to select your output type: 0-20 or 4-20 mA.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.5.3 step 3, or return to Measurement mode by pressing
the and keys simultaneously (escape).
7.5.3 Selecting pH/ORP value at 0(4)mA
This parameter lets you choose a pH or ORP value at which the controller
output will be 0(4) mA. NOTE: If you are in the pH mode, ORP parameters
are blocked out. If you are in the ORP mode, pH parameters are blocked out.
1) Follow directions in 7.5.1 to enter Current Output mode.
2) Press the ENTER key. Scroll with the or keys until the upper display shows a pH or ORP value
and the lower display shows “r.0(4)”.
3) Press the or key to select pH or ORP value to be equivalent to 0(4) mA.
4) Press the ENTER key to confirm your selection.
5) Press the ENTER key to return to Advanced Set-up mode, or return to Measurement mode by pressing
the and keys simultaneously (escape).
7.5.4 Selecting pH/ORP value at 20mA
This parameter lets you choose a pH or ORP value at which the controller output will be 20mA.
NOTE: If you are in the pH mode, ORP parameters are blocked out. If you are in the ORP mode, pH
parameters are blocked out.
1) Follow directions in 7.5.1 to enter Current Output mode.
2) Press the ENTER key. Scroll with the or key until the upper display shows a pH or ORP value and
the lower display shows “r.20”.
3) Press the or key to select pH or ORP value to be equivalent to 20 mA.
4) Press the ENTER key to confirm your selection.
5) Press the ENTER key to return to Advanced Set-up mode, or return to Measurement mode by pressing
the and keys simultaneously (escape).
rng
SETUP
HOLD
4-20
out
SETUP
HOLD
SETUP
0
pH
r.4
HOLD
14
pH
r.20
SETUP
HOLD