6.6 Cell Constant Setting
You can select up to three types of cell constant in the setup mode: - K=1.0, K=0.1
or K=10.
From the measurement mode,
1. Press the ENT key to enter the setup mode. Use
the ▲ or ▼ keys to scroll till LCD displays ‘SEt’ in
the upper display; and ‘CEL’ in the lower display.
2. Press the ENT key to enter the cell constant setup
page.
3. Use the ▲ or ▼ keys to select the desired cell
constant and press the ENT key to confirm the
selection.
4. Press S and T keys together, to return to the
measurement mode.
6.7 Line resistance Setting
Line resistance of an electrode cable constitutes to error in measurement of high
conductivity values as in range 6. This error can be compensated by inputting the
electrode cable line resistance value in the setup mode.
From Range 6 measurement mode,
1. Press the ENT key to enter the setup mode. Use
the ▲ or ▼ keys to scroll till LCD displays ‘SEt’ in
the upper display; and ‘Lnr’ in the lower display.
2. Press the ENT key to enter the line resistance
setup page.
3. Use the ▲ or ▼ keys to set the line resistance
value (0 to 50 ohms) and press the ENT key to
confirm the setting.
4. Press S and T keys together, to return to the
measurement mode.
NOTE: The line resistance setup mode can be accessed
only from range 6 measurement mode.