8.9 Controller Reset – P7.0
Program 7 gives an option to reset the controller to factory default values.
• Conductivity range remains unchanged based on last selected range
• All calibration data and conductivity calibration factor are reset
• SP 1 is reset to 10 % of full scale
• SP 2 is reset to 90 % of full scale
• Hysteresis for both set points is reset to 5 % of full scale
• Temperature compensation mode remains unchanged (ATC or MTC)
• In MTC mode, temperature value is reset to 25
o
C
• Temperature Coefficient is reset to 2.1%
• Normalization Temperature is reset to 25.0
o
C
NOTE: ATC temperature offset is NOT reset.
P7. 0
RsT
n0
yeS
P7. 1
Press ▲and ▼keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P7.0: Controller reset
1. Key in the password “022” as per procedure in Section 5.
2. Press ▲key until screen displays P7.0 and RST. Press ENTER.
3. Screen will scroll P7.1, then ‘No’.
4. Press ▲or ▼keys to toggle between YES and NO. Press ENTER to
confirm selection.
5. Press ▲and ▼keys together (ESCAPE) to return to P7.0. Press ▲and
▼keys together again, to return to measurement mode.