8.9 Controller Reset – P7.0
Program 7 gives an option to reset the controller to factory default values.
Conductivity range remains unchanged based on last selected range
All calibration data and conductivity calibration factor are reset
SP 1 is reset to 10 % of full scale
SP 2 is reset to 90 % of full scale
Hysteresis for both set points is reset to 5 % of full scale
Temperature compensation mode remains unchanged (ATC or MTC)
In MTC mode, temperature value is reset to 25
o
C
Temperature Coefficient is reset to 2.1%
Normalization Temperature is reset to 25.0
o
C
NOTE: ATC temperature offset is NOT reset.
P7. 0
RsT
n0
yeS
P7. 1
Press and keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P7.0: Controller reset
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P7.0 and RST. Press ENTER.
3. Screen will scroll P7.1, then ‘No’.
4. Press or keys to toggle between YES and NO. Press ENTER to
confirm selection.
5. Press and keys together (ESCAPE) to return to P7.0. Press and
keys together again, to return to measurement mode.