Measurement: Cyanuric Acid C401/301/201/102/101
16
Add content of one
Cyanuric Acid sachet
Shake for
1 minute
9. Gently push the vial back fully into the sample well of the instrument,
whilst again aligning the mark on the vial with the mark on the
meter.
10. Wait for two minutes and then press the READ/ENTER key within
five minutes from when the sachet was added to the vial.
11. The meter will flash [CYNA] whilst the measurement is taken. It will
then display the Cyanuric Acid concentration in parts per million
(ppm).
Cyanuric
Acid
Sample
READ
ENTER
ppm